- Product Name: Coupling Decoupling Network (CDN)
- Compliant Test Standards: IEC / EN 61000-4-6
- Application: Eight wire unscreened, unbalanced cables
- Frequency Range: 150 kHz to 230 MHz
- RF Input Voltage: 40 V (Max)
- RF Input Connector: 50 Ω BNC (Female)
- Voltage Rating: 310 V AC / 440 V DC (Line to Ground)
- Current Rating: 5 Amp (Max)
- AE and EUT Connections: 2 mm shrouded banana sockets
- Common Mode Impedance:
- 150 kHz – 26 MHz: 150Ω ± 20Ω
- 26 MHz – 80 MHz: 150Ω + 60Ω / – 45Ω
- 80 MHz – 230 MHz: 150Ω + 60Ω / – 60Ω
- Voltage Division Factor: 9.5 dB +3 / -1
- Decoupling of Common Mode Disturbance: ≥ 45dB (EUT/AE)
- Dimensions: 6 x 6 x 13 inches or 15.2 x 15.2 x 33 cm
- Weight: 5 lbs. or 2.3 kg
- Accessories Available from Com-Power for setting test levels and running the test:
- ADA-AF8E shorting adapters
- ADA-515-2 150 Ω to 50 Ω adapters
- TEP-050 50 Ω Terminator
- ATTN-6-100W Power Attenuator
- DCU-300-100W Directional Coupler
- ASC series Power Amplifiers
[:]